Gastrointestinal infection (GI) - November 01, 2017
Multisite evaluation of the BD MAX™ Extended Enteric Bacterial Panel
Multisite Evaluation of the BD MAX™ Extended Enteric Bacterial Panel for Detection of Yersinia enterocolitica, Enterotoxigenic Escherichia coli, Vibrio, and Plesiomonas shigelloides from Stool Specimens
Patricia J Simner 1, Margret Oethinger 2, Kathleen A Stellrecht 3, Dylan R Pillai 4, Ram Yogev 5, Helene Leblond 6, Joel Mortensen 7
November 2017
Abstract:
The purpose of this study was to perform a multisite evaluation to establish the performance characteristics of the BD MAX™ extended enteric bacterial panel (xEBP) assay directly from unpreserved or Cary-Blair-preserved stool specimens for the detection of Yersinia enterocolitica, enterotoxigenic Escherichia coli (ETEC), Vibrio, and Plesiomonas shigelloides.
The study included prospective, retrospective, and prepared contrived specimens from 6 clinical sites. BD MAX™ xEBP results were compared to the reference method, which included standard culture techniques coupled with alternate PCR and sequencing, except for ETEC, for which the reference method was two alternate PCRs and sequencing.
Alternate PCR was also used to confirm the historical results for the retrospective specimens and for discrepant result analysis. A total of 2,410 unformed, deidentified stool specimens were collected.
The prevalence in the prospective samples as defined by the reference method was 1.2% ETEC, 0.1% Vibrio, 0% Y. enterocolitica, and 0% P. shigelloides Compared to the reference method, the positive percent agreement (PPA) (95% confidence interval [CI]), negative percent agreement (NPA) (95% CI), and kappa coefficient (95% CI) for the BD MAX™ xEBP assay for all specimens combined were as follows: ETEC, 97.6% (87.4 to 99.6), 99.8% (99.5 to 99.9), and 0.93 (0.87 to 0.99); Vibrio, 100% (96.4 to 100), 99.7% (99.4 to 99.8), and 0.96 (0.93 to 0.99); Y. enterocolitica, 99.0% (94.8 to 99.8), 99.9% (99.8 to 99.9), and 0.99 (0.98 to 1); P. shigelloides, 100% (96.4 to 100), 99.8% (99.5 to 99.9), and 0.98 (0.95 to 1), respectively.
In this multicenter study, the BD MAX™ xEBP showed a high correlation (kappa, 0.97; 95% CI, 0.95 to 0.98) with the conventional methods for the detection of ETEC, Vibrio, Y. enterocolitica, and P. shigelloides in stool specimens from patients suspected of acute gastroenteritis, enteritis, or colitis.